The LogTag® UTRED30-16 Temperature Recorder measures and stores up to 16,000 temperature readings per channel, a total of 32,000 temperature readings.
It features a large easy-to-read display and is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 29 days.
The alarm is triggered if temperature readings are outside pre-set alarm limits which is indicated by a large X on the display and, if configured, the audible alarm will sound.
UTRED30-16
-40°C to +99°C (-40°F to +210°F)
-30°C to +60°C (-22°F to +140°F)
0 – 99 % RH (non-condensing)
-10°C to +40°C (32°F to +104°F)
Better than 0.1°C (0.2°F) for -30°C to +60°C (-22°F to +140°F).
Better than 0.2°C (0.4°F) for -40°C to -30°C (-40°F to -22°F).
Better than 0.5°C (0.9°F) for +60°C to +99°C (+140°F to +210°F).
LogTag Analyzer currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F), typically ±0.3°C (0.6°F)
Better than ±0.8°C (±1.4°F) for -40°C to -20°C (-40°F to -4°F), typically ±0.5°C (0.9°F)
Better than ±0.8°C (±1.4°F) for +40°C to +70°C (+104°F to +158°F), typically ±0.7°C (1.3°F)
Better than ±1.2°C (±2.2°F) for +70°C to +99°C (+158°F to +210°F), typically ±1°C (1.8°F)
Up to 2
All ST100 precision thermistor external probes can be interchangeably connected to the UTRED30-16.
Please refer to the Accessories section for compatible External Probes. Different probe models can be used for each channel. Not compatible with ST10 sensors.
Typically less than 2 minutes(T90) in moving air (1m/s) for ST100S type
32,000 total log memory. Provides 16,000 temperature readings per channel.
Day summary statistics memory (for display on LCD): up to 30 days of Min/Max/duration and alarm trigger statistics.
Configurable from 30 seconds to 18 hours (Typically 5 minute logging)
Up to 6 individually configurable alarms for each channel
Possible to set alarms separately for Channel 1 and Channel 2
Push button start or specific date & time. Optional start delay
Typically with full memory (32,000 readings) in less than 20 seconds
IP61 (when vertically mounted or hung with USB dust-cap in place)
5V via micro USB socket or 2 x 1.5V AAA batteries
CR2032 LiMnO2 coin cell (backup logging function only)
AAA battery up to 12 months – if unit is operated without power supply
Built-in real time clock
Typical accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day)
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
USB 2.0 / Micro USB Socket
Standard: 1.5m (4’11”)
Extended: 3m (9’10”) (recommended maximum)
PTFE (FDA food contact rated) coaxial
141mm(H) x 72mm(W) x 18mm(T)
156g with AAA batteries, (132g – excluding probe and AAA batteries)
Polycarbonate
UTRED30-16 Brochure
UTRED30-16 Quick Start Guide
UTRED30-16 User Manual